User manual

Remote Control Commands
R&S
®
Scope Rider RTH
180User Manual 1326.1578.02 ─ 04
Parameters:
<Condition> LONGer | SHORter | EQUal | NEQual | WITHin | OUTSide
LONGer | SHORter
Triggers on pulses shorter or longer than a runt width that is set
using TRIGger:SLEW:RATE.
EQUal | NEQual
Triggers on pulses equal or not equal a given runt width that is
set using TRIGger:SLEW:RATE. In addition, a tolerance can be
set around the specified width using TRIGger:RUNT:DELTa.
WITHin | OUTSide
Triggers on pulses inside or outside a given range. The range is
set using ... and ....
*RST:
LONGer
TRIGger:SLEW:RATE <Duration>
Sets the slew rate for comparison ranges EQUal, UNEQual, SHORter, and LONGer.
See TRIGger:SLEW:RANGe.
Parameters:
<Duration> Range: 800E-12 to 10000
Increment: 100E-9
*RST: 5E-9
Default unit: s
TRIGger:SLEW:DELTa <Tolerance>
Sets a range Δt to the specified slew rate, which is defined using TRIGger:SLEW:
RATE.
Parameters:
<Tolerance> Range: 0 to 5000
Increment: 500E-12
*RST: 0
Default unit: s
12.3.5.9 Data2Clock Trigger
See also Chapter 2.5.10, "Data2Clock Trigger (R&S RTH-K19)", on page 64.
TRIGger:DATatoclock:DSOurce[:VALue]..........................................................................181
TRIGger:DATatoclock:CSOurce[:VALue]..........................................................................181
TRIGger:DATatoclock:CSOurce:EDGE............................................................................ 181
TRIGger:DATatoclock:CONDition.................................................................................... 181
TRIGger:DATatoclock:HTIMe..........................................................................................181
TRIGger:DATatoclock:STIMe.......................................................................................... 182
Waveform Setup