User manual

Remote Control Commands
R&S
®
Scope Rider RTH
228User Manual 1326.1578.02 ─ 04
Suffix:
<m>
.
1..3
1 = all logic channels D0 to D7
2 = group D0 to D3
3= group D4 to D7
Return values:
<Level> Range: -10 to 10
Increment: 1E-3
*RST: 0
Default unit: V
Usage: Query only
LOGic:GROup<m>:HYSTeresis <Hysteresis>
Hysteresis avoids the change of signal states due to noise oscillation around the
threshold level. Set a small hysteresis for clean signals, and large hysteresis for noisy
signals.
Suffix:
<m>
.
1..3
1 = all logic channels D0 to D7
2 = group D0 to D3
3= group D4 to D7
The suffix only takes effect if LOGic:THCoupling is OFF.
Parameters:
<Hysteresis> SMALl | MEDium | LARGe
*RST: MEDium
POD:STATe?
Returns the connection state of the logic probe.
Return values:
<PODConnected> ON | OFF
*RST: OFF
Usage: Query only
Logic Analyzer (R&S RTH-B1 MSO)