User manual

Waveform Analysis
R&S
®
Scope Rider RTH
83User Manual 1326.1578.02 ─ 04
Remote command:
CURSor:FUNCtion on page 191
Source
Defines the source on which the cursor measurement is performed. The source can be
any active analog or digital input signal, math waveform, or bus (requires option).
The source setting is not available for the cursor type "Vertical", and for measurements
in the XY-diagram.
Remote command:
CURSor:SOURce on page 191
Show Position
Shows the position values of the vertical cursor lines t1 and t2 in the measurement
results. The setting is only available for the vertical cursor type.
Meas Type 1 / Meas Type 2
Set the automatic measurements to be performed on the source waveform between
the cursor lines. The setting is only available for the "Measure" cursor type.
All automatic measurements that need only one source are available. Delay, phase,
and power measurements are not provided for cursor measurements, they require two
sources.
For a description of the measurement types, see Chapter 3.2.2, "Measurement Types",
on page 74.
Remote command:
CURSor:MEASurement<m>:TYPE on page 192
Track Scaling
If enabled, the position of the cursor lines is adjusted if the vertical or horizontal scales
are changed. The cursor lines keep their relative position to the waveform.
If disabled, the cursor lines remain on their position on the display if the scaling is
changed.
Remote command:
CURSor:SCPLing on page 192
Coupling
Couples the cursor lines so that the distance between the two lines remains the same
if one cursor is moved.
Remote command:
CURSor:COUPling on page 191
Set to Screen
Sets the cursors to a default position on the screen. This is helpful if the cursors have
disappeared from the display or need to be moved for a larger distance.
Cursor Measurements