5.1.3 Running Voltmeter Measurements
1. Select the "Vertical" menu and adjust the "Probe Setting" for all channels to be
measured.
See also "Probe Setting" on page 42.
2. If the 4-voltmeter-view is shown, switch to the one-voltmeter-view, see "Displaying
voltmeters" on page 102.
3. Select the channel to be measured (no. 8 in Figure 5-1).
4. Select the measurement type (no. 9).
5. Activate the voltmeter (no. 12)
6. To adjust the measurement range manually, press the RANGE keys.
7. If you want to get relative result values:
a) Activate "Relative" (no. 11).
b) To set a user-defined reference value, open the "Meter" menu, activate "Refer-
ence", and enter the value in "Reference Manual".
Alternatively, you can select the voltmeter and adjust the meter settings in the "Meter"
menu.