User manual

Logic Analyzer (R&S RTH-B1 MSO)
R&S
®
Scope Rider RTH
137User Manual 1326.1578.02 ─ 04
Threshold
Logic 0 Logic 1 Logic 0
Hysteresis
Remote command:
LOGic:GROup<m>:HYSTeresis on page 228
Deskew
Enables deskewing.
Deskew compensates delays that are known from the circuit specifics or caused by the
different length of cables. The skew between the probe boxes of the digital channels
and the probe connectors of the analog channels is automatically aligned by the instru-
ment.
Select Channel / Deskew
You can set the deskew for all channels of a logic probe at once, or for each logic
channel separately.
Select the channel and enter the deskew value in "Deskew".
Set all Deskews to Zero
Resets all deskew values to zero.
Logic Analyzer Settings