Selftest
The self-test checks the hardware of the instrument. It is intended for service tasks.
USB/LAN
See Chapter 11.1, "LAN Connection", on page 153 and Chapter 11.2, "USB connec-
tion", on page 155.
Wireless LAN
See Chapter 11.3, "Wireless LAN Connection (Option R&S RTH-K200)", on page 155.
Time and Date
See Chapter 10.3, "Setting the Date, Time and Language", on page 146.
User Interface Language
See Chapter 10.3, "Setting the Date, Time and Language", on page 146.
Options
See Chapter 10.4, "Options", on page 147
Maintenance
The "Device Info" tab provides service information for your R&S RTH. If you need sup-
port, you may be asked to provide this information. Here you can also read the Open
Source Acknowledgment, which provides verbatim license text of open source soft-
ware that is used in the instrument's firmware.
The "Service" tab allows the service personnel to enter a password that activates fur-
ther service functions.
Firmware Update
See Chapter 10.5, "Updating the Firmware", on page 150.
10.2 Selfalignment
Access:
or "Setup" menu > "Selfalignment"
The self-alignment aligns the data from several input channels vertically and horizon-
tally to synchronize the timebases, amplitudes and positions.
The R&S RTH1002 provides an additional, separate self-alignment of the meter inputs.
Recommendation on performing the self-alignment:
When putting the instrument into operation for the first time
After a firmware update
Once a week
When major temperature changes occur (> 5°)
Selfalignment