User manual

Remote Control Commands
R&S
®
Scope Rider RTH
177User Manual 1326.1578.02 ─ 04
TRIGger:STATe:COMBination <ChCombination>
Sets the logical combination for all active channels. The required state of each channel
is defined by TRIGger:PATTern:STATe[:CHANnel<1..22>].
Parameters:
<ChCombination> AND | OR
*RST: AND
TRIGger:STATe:CSOurce[:VALue] <Source>
Sets the input channel of the clock signal.
Parameters:
<Source> C1 | C2 | C3 | C4 | D0 | D1 | D2 | D3 | D4 | D5 | D6 | D7
*RST: C1
TRIGger:STATe:CSOurce:EDGE
<Slope>
Sets the edge of the clock at which the instrument checks the signal states.
Parameters:
<Slope> POSitive | NEGative | EITHer
*RST: POSitive
12.3.5.7 Runt Trigger
See also Chapter 2.5.8, "Runt Trigger (R&S RTH-K19)", on page 60.
TRIGger:LEVel<m>:RUNT:LOWer...................................................................................177
TRIGger:LEVel<m>:RUNT:UPPer................................................................................... 177
TRIGger:RUNT:POLarity................................................................................................ 178
TRIGger:RUNT:RANGe..................................................................................................178
TRIGger:RUNT:WIDTh...................................................................................................178
TRIGger:RUNT:DELTa...................................................................................................178
TRIGger:LEVel<m>:RUNT:LOWer <LowerLevel>
TRIGger:LEVel<m>:RUNT:UPPer <UpperLevel>
Set the upper and lower levels that limit the runt.
Suffix:
<m>
.
1..22
1..4: Indicates the trigger source: analog channel 1 to 4
5..22: not available
Parameters:
<LowerLevel>
<UpperLevel>
Range: -10 to 10
Increment: 1E-3
*RST: 0
Default unit: V
Waveform Setup