User manual

Remote Control Commands
R&S
®
Scope Rider RTH
181User Manual 1326.1578.02 ─ 04
TRIGger:DATatoclock:DSOurce[:VALue] <Source>
Selects the input channel of the data signal.
Parameters:
<Source> C1 | C2 | C3 | C4 | D0 | D1 | D2 | D3 | D4 | D5 | D6 | D7
*RST: C1
TRIGger:DATatoclock:CSOurce[:VALue] <Source>
Selects the input channel of the clock signal.
Parameters:
<Source> C1 | C2 | C3 | C4 | D0 | D1 | D2 | D3 | D4 | D5 | D6 | D7
*RST: C1
TRIGger:DATatoclock:CSOurce:EDGE <Slope>
Sets the edge of the clock signal: rising (POSitive), falling (NEGative), or both
edges (EITHer). The time reference point for the setup and hold time is the crossing
point of the clock edge and the trigger level.
Parameters:
<Slope> POSitive | NEGative | EITHer
*RST: POSitive
TRIGger:DATatoclock:CONDition <TrigCondition>
Selects how a violation of the setup and hold time is handled.
Parameters:
<TrigCondition> VIOLation | OK
VIOLation
Triggers on a violation of the setup or hold time
OK
Triggers if setup and hold time keep the limits.
*RST: VIOLation
TRIGger:DATatoclock:HTIMe <HoldTime>
Sets the minimum time after the clock edge while the data signal must stay steady.
The hold time can be negative. In this case, the setup time has to be positive. The
setup time is defined by TRIGger:DATatoclock:STIMe.
Waveform Setup