184User Manual 1326.1578.02 ─ 04
TRIGger:TIMeout:RANGe <TimeoutMode>
Selects the relation of the signal level to the trigger level, which is specified with
TRIGger:LEVel<m>:VALue
Parameters:
<TimeoutMode> HIGH | LOW | EITHer
*RST: HIGH
TRIGger:TIMeout:TIME <Time>
Sets the time limit for the timeout at which the instrument triggers.
Parameters:
<Time> Range: 100E-12 to 10000
Increment: 100E-9
*RST: 5E-9
Default unit: s
12.3.5.12 Window Trigger
TRIGger:LEVel<m>:WINDow:LOWer............................................................................... 184
TRIGger:LEVel<m>:WINDow:UPPer................................................................................184
TRIGger:WINDow:TIME..................................................................................................184
TRIGger:WINDow:RANGe.............................................................................................. 185
TRIGger:WINDow:WIDTh............................................................................................... 185
TRIGger:WINDow:DELTa............................................................................................... 186
TRIGger:LEVel<m>:WINDow:LOWer <Level>
TRIGger:LEVel<m>:WINDow:UPPer <Level>
Set the upper and lower voltage thresholds for the window trigger, respectively. The
trigger levels are the vertical window limits.
Suffix:
<m>
.
1..22
1..4: Indicates the trigger source: analog channel 1 to 4
5..22: not available
Parameters:
<Level> Range: -10 to 10
Increment: 1E-3
*RST: 0
Default unit: V
TRIGger:WINDow:TIME <Condition>
Selects how the time limit of the window is defined.
Waveform Setup