User manual

Remote Control Commands
R&S
®
Scope Rider RTH
185User Manual 1326.1578.02 ─ 04
Parameters:
<Condition> LONGer | SHORter | EQUal | NEQual | WITHin | OUTSide
LONGer | SHORter
Triggers if the signal crosses the upper or lower level after/
before the time "Width" defined by TRIGger:WINDow:WIDTh
EQUal | NEQual
Triggers if the signal stays inside or outside the vertical window
limits for a time equal/uneuqal to "Width" "±Tolerance" defined
by TRIGger:WINDow:WIDTh and TRIGger:WINDow:DELTa.
WITHin
Triggers if the signal stays inside or outside the vertical window
limits for a time "Min Width" AND "Max Width".
OUTSide
Triggers if the signal stays inside or outside the vertical window
limits for a time < "Min Width" OR > "Max Width".
*RST:
LONGer
TRIGger:WINDow:RANGe <LevelRangeMode>
Selects how the signal run is compared with the window:
Parameters:
<LevelRangeMode> ENTer | EXIT | WITHin | OUTSide
ENTer | EXIT
Triggers when the signal crosses the upper or lower level and
thus enters/leaves the window made up of these two levels
which are defined by TRIGger:LEVel<m>:WINDow:UPPer
and TRIGger:LEVel<m>:WINDow:LOWer
WITHin | OUTSide
Triggers if the signal stays between/above the upper and lower
level for a specified time. The time is defined by TRIGger:
WINDow:TIME
*RST: ENTer
TRIGger:WINDow:WIDTh <Duration>
Sets the width for comparison ranges LONGer, SHORter, EQUal, NEQual.
See TRIGger:WINDow:TIME on page 184.
Parameters:
<Duration> Range: 800E-12 to 10000
Increment: 100E-9
*RST: 5E-9
Default unit: s
Waveform Setup