User manual

Remote Control Commands
R&S
®
Scope Rider RTH
189User Manual 1326.1578.02 ─ 04
The sources can be any active input signal, math or reference waveform. Available
source waveforms depend on the measurement type, see Chapter 3.2.2, "Measure-
ment Types", on page 74.
Suffix:
<m>
.
1..4
Parameters:
<Source> C1 | C2 | C3 | C4 | M1 | R | D0 | D1 | D2 | D3 | D4 | D5 | D6 | D7
<Source2> C1 | C2 | C3 | C4 | M1 | R | D0 | D1 | D2 | D3 | D4 | D5 | D6 | D7
MEASurement<m>:TYPE
<Type>
Selects the measurement type. For a detailed description, see Chapter 3.2.2, "Mea-
surement Types", on page 74.
Suffix:
<m>
.
1..4
Parameters:
<Type> PERiod | FREQuency | RTIMe | FTIMe | PPULse | NPULse |
PDCYcle | NDCYcle | DELay | PHASe | MEAN | RMS | CRESt |
STDDev | MINimum | MAXimum | PKPK | BASelevel |
TOPLevel | AMPLitude | OVRShoot | PREShoot | AC | DC |
ACDC | PPCount | NPCount | RECount | FECount | PWRP |
PWRS | PWRQ | PWRFactor
*RST: MINimum
RTIMe Rise time
PREShoot Preshoot
FTIMe Fall time
PPCount Positive pulse count
PPULse Positive pulse width
NPCount Negative pulse count
NPULse Negative pulse width
RECount Rising edge count
PDCYcle Positive duty cycle
FECount Falling edge count
NDCYcle Negative duty cycle
PWRP Active power
STDDev Standard deviation
PWRS Apparent power
PKPK Peak to peak
PWRQ Reactive power
OVRShoot Overshoot
PWRFactor Power factor
MEASurement<m>:AOFF
Disables all active measurements.
Suffix:
<m>
.
1..4
The suffix is irrelevant.
Usage: Event
Waveform Analysis