User manual

Remote Control Commands
R&S
®
Scope Rider RTH
200User Manual 1326.1578.02 ─ 04
Parameters:
<AutoRepeat> ON | OFF
*RST: OFF
CHANnel<m>:HISTory:TSABsolute?
Returns the abolsute daytime of the current segment (CHANnel<m>:HISTory:
CURRent).
Return values:
<TimeStampAbsTime>String containing the time and unit.
Usage: Query only
CHANnel<m>:HISTory:TSDate?
Returns the date of the current segment (CHANnel<m>:HISTory:CURRent).
Return values:
<TimeStampAbsData>String with date of the current acquisition (absolute time)
Usage: Query only
CHANnel<m>:HISTory:TSRelative?
Returns the relative time of the current segment - the time difference to the newest
segment (index = 0).
See also CHANnel<m>:HISTory:CURRent.
Return values:
<TimeStampRel> String containing the relative time in seconds.
Usage: Query only
12.5 Mask Testing
12.5.1 Mask Definition
The suffix <m> selects the mask channel for which the command is executed. The
number of channels depends on the instrument type. The R&S RTH1004 has 5 mask
channels, the suffix values are 1 | 2 | 3 | 4 | 5. The R&S RTH1002 has 3 mask chan-
nels, the suffix values are 1 | 2 | 5. Suffix 5 is used for the mask on a math waveform.
MASK:CHANnel<m>:STATe........................................................................................... 201
MASK:CHANnel<m>:PROPerties:XWIDth........................................................................ 201
MASK:CHANnel<m>:PROPerties:YWIDth........................................................................ 201
MASK:CHANnel<m>:CREatemask.................................................................................. 201
Mask Testing