MASK:CHANnel<m>:STATe <State>
Turns the selected mask on or off.
Suffix:
<m>
.
1|2|5 (RTH1002) , 1..5 (RTH1004)
Parameters:
<State> ON | OFF
*RST: OFF
MASK:CHANnel<m>:PROPerties:XWIDth <WidthX>
Changes the width of the selected mask in horizontal direction.
Suffix:
<m>
.
1|2|5 (RTH1002) , 1..5 (RTH1004)
Parameters:
<WidthX> Range: 0 to 10
Increment: 0.01
*RST: 0
Default unit: div
MASK:CHANnel<m>:PROPerties:YWIDth <WidthY>
Changes the width of the selected mask in vertical direction.
Suffix:
<m>
.
1|2|5 (RTH1002) , 1..5 (RTH1004)
Parameters:
<WidthY> Range: 0 to 8
Increment: 0.01
*RST: 0.1
Default unit: div
MASK:CHANnel<m>:CREatemask
Creates a mask from the envelope waveform of the selected waveform with the
defined width i x and y direction.
Suffix:
<m>
.
1|2|5 (RTH1002) , 1..5 (RTH1004)
Usage: Event
Mask Testing