User manual

Remote Control Commands
R&S
®
Scope Rider RTH
223User Manual 1326.1578.02 ─ 04
Parameters:
<ChipSelectPolarity> ACTLow | ACTHigh
*RST: ACTHigh
BUS:SPI:TECHnology <ThresholdType>
Sets the threshold values of all SPI lines as defined for various signal technologies.
Parameters:
<ThresholdType> TTL | ECL | CMOS | USER
USER: Set a user-defined value for each line using the
BUS:SPI:<line>:THReshold commands.
*RST: TTL
Usage: SCPI confirmed
BUS:SPI:SCLK:THReshold <ThresUserValue>
BUS:SPI:SSEL:THReshold <ThresUserValue>
BUS:SPI:MOSI:THReshold <ThresUserValue>
BUS:SPI:MISO:THReshold <ThresUserValue>
Set a user-defined value for the corresponding line if BUS:SPI:TECHnology is set to
USER.
Parameters:
<ThresUserValue> Range: -8 to 8
Increment: 1E-3
*RST: 1.4
Default unit: V
BUS:SPI:WSIZe <WordLength>
Sets the number of bits in a word.
Parameters:
<WordLength> WL4Bit | WL8Bit | WL12bit | WL16bit | WL20bit | WL24bit |
WL28bit | WL32bit
*RST: WL8Bit
12.8.4 UART/RS-232/RS-422/RS-485 (Option R&S RTH-K2)
12.8.4.1 UART Configuration
BUS:UART:SOURce...................................................................................................... 224
BUS:UART:POLarity...................................................................................................... 224
BUS:UART:THReshold...................................................................................................224
BUS:UART:STDBitrate................................................................................................... 224
Protocol Analysis