Parameters:
<PersistenceType> OFF | TIME | INF
OFF
Deactivates persistence.
TIME
Data points remain on the screen for the duration defined with
DISPlay:PERSistence:TIME.
INF
Data points remain on the screen infinitely until persistence is
set to OFF.
*RST: OFF
DISPlay:PERSistence:TIME
<PersistenceTime>
Sets a user-defined persistence time. The command takes effect if DISPlay:
PERSistence[:TYPE] is set to TIMe.
Parameters:
<PersistenceTime> Range: 0.05 to 10
Increment: 0.01
*RST: 0.05
Default unit: s
DISPlay:CONTrast <ContrastMode>
If enabled, the waveforms are displayed in black color on white background.
Parameters:
<ContrastMode> ON | OFF
*RST: OFF
DIAGnostic:SERVice:LCD:BRIGthness <LCDBrightness>
Changes the brightness of the touchscreen.
Parameters:
<LCDBrightness> LOW | MEDium | HIGH
*RST: MEDium