User manual

Waveform Setup
R&S
®
Scope Rider RTH
50User Manual 1326.1578.02 ─ 04
Remote command:
TRIGger:HOLDoff:EVENts on page 170
Min Time / Max Time ← Holdoff Mode
Set the time limits for random holdoff time. For each acquisition, the instrument selects
a new random holdoff time from the specified range.
Remote command:
TRIGger:HOLDoff:MIN on page 170
TRIGger:HOLDoff:MAX on page 170
Noise Reject
Enables a hysteresis to avoid unwanted trigger events caused by noise oscillation
around the trigger level.
Remote command:
TRIGger:MNR on page 171
2.5.2 Edge Trigger
The edge trigger is the most common trigger type. The trigger occurs when the signal
from the trigger source passes the trigger level in the specified direction (slope).
Description of settings
Figure 2-2: Edge trigger
Slope
Sets the edge direction for the trigger. You can trigger on:
Rising edge, that is a positive voltage change
Falling edge, that is a negative voltage change
Rising and falling edge
Remote command:
TRIGger:EDGE:SLOPe on page 171
Trigger