User manual

Waveform Setup
R&S
®
Scope Rider RTH
60User Manual 1326.1578.02 ─ 04
Clock Slope
Sets the edge of the clock at which the instrument checks the signal states: at the ris-
ing edge, the falling adge, or at both edges.
Remote command:
TRIGger:STATe:CSOurce:EDGE on page 177
Set Pattern
Defines the pattern: the states of the input channels and their logical combination. If
R&S RTH-B1 is installed, active digital channels are also included in the pattern defini-
tion.
For details, see Chapter 2.5.6, "Pattern Trigger (R&S RTH-K19)", on page 57.
Remote command:
TRIGger:STATe:CHANnel<1..22> on page 176
TRIGger:STATe:COMBination on page 177
2.5.8 Runt Trigger (R&S RTH-K19)
A runt is a pulse lower than normal in amplitude. The amplitude crosses the first level
twice in succession without crossing the second one. In addition to the upper and lower
levels, you can define a time limit for the runt in the same way as for width triggers. For
example, the runt trigger can detect signal parts remaining below a specified threshold
amplitude because I/O ports are in undefined state.
UL
L L
Figure 2-13: Runt trigger without time limits
SETUP key > "Trigger Type" = "Runt"
Trigger