User manual

Waveform Setup
R&S
®
Scope Rider RTH
63User Manual 1326.1578.02 ─ 04
Description of settings
Upper Trigger Level / Lower Trigger Level
Set the upper and lower voltage thresholds for the slew rate trigger. The time measure-
ment starts when the signal crosses the first trigger level - the upper or lower level
depending on the selected slope - and stops when the signal crosses the second level.
Remote command:
TRIGger:LEVel<m>:SLEW:UPPer on page 179
TRIGger:LEVel<m>:SLEW:LOWer on page 179
Slope
Sets the edge direction for the trigger. You can trigger on:
Rising edge, that is a positive voltage change
Falling edge, that is a negative voltage change
Rising and falling edge
Remote command:
TRIGger:SLEW:SLOPe on page 179
Range
Defines the time limits of the slew rate. The comparisons are the same as for the width
trigger, see "Range" on page 52.
Remote command:
TRIGger:SLEW:RANGe on page 179
Time
Sets the slew rate for comparisons equal, unequal, shorter, and longer.
Remote command:
TRIGger:SLEW:RATE on page 180