Datasheet

Version 08.00, March 2016
Rohde & Schwarz R&S
®
RTH1002, R&S
®
RTH1004 Scope Rider 15
Serial pattern triggers on serial data pattern up to 32 bit clocked by one input channel; pattern bits
may be high (H), low (L) or don’t care (X); clock edge slope may be positive, negative
or either (requires R&S
®
RTH-K19 option)
max. data rate < 250 Mbps
Timeout triggers when signal stays high, low or unchanged for a specified period of time
(requires R&S
®
RTH-K19 option)
timeout 200 ps to 5000 s (CH1, CH2, CH3, CH4)
800 ps to 5000 s (D0 to D7)
Interval triggers when time between two consecutive edges of same slope (positive or
negative) is shorter, longer inside or outside a specified range (requires R&S
®
RTH-K19
option)
interval time 200 ps to 5000 s (CH1, CH2, CH3, CH4)
800 ps to 5000 s (D0 to D7)
Protocol see R&S
®
RTH-K1 and R&S
®
RTH-K2 options
Waveform measurements
Automatic measurements on total number of active measurements 4
sources logic channels from D0 to D7
time based measurements period, frequency, positive pulse width,
negative pulse width, positive duty cycle,
negative duty cycle, delay, phase
amplitude based measurements mean value
count based measurements count positive pulses, count negative
pulses, count rising edges, count falling
edges
Cursor measurements sources logic channels from D0 to D7
vertical 2 cursors showing time, time difference
and inverse time difference (frequency)
tracking vertical cursor additionally showing logic
level and logic level difference of selected
channel
measure defines gate for automatic measurements