Datasheet

Version 08.00, March 2016
Rohde & Schwarz R&S
®
RTH1002, R&S
®
RTH1004 Scope Rider 17
R&S
®
RTH-K2
UART/RS-232/RS-422/RS-485 serial triggering and decoding
Protocol configuration bit rate 300 bps to 20 Mbps
signal polarity idle low, idle high
number of bits 5 bit to 9 bit
bit order LSB first, MSB first
parity odd, even, none
stop bits 1, 1.5 or 2
end of packet timeout, none
auto threshold setup assisted threshold configuration for
UART triggering and decoding (software)
Trigger source any input channel or logical channel
trigger event setup start bit, packet start, data, parity error,
stop error, break condition
data setup data pattern (hex, decimal, octal, binary or
ASCII); condition =, ; >, <; offset within
packet in range 0 to 4095 words
Decode source any input channel, logical channel
display type decoded bus, tabulated list
color coding start, data payload, parity, stop, start
error, parity error, stop error
data format hex, decimal, octal, binary, ASCII
R&S
®
RTH-K15
History and segmented memory
Memory segmentation function additional memory segments for the acquisition
number of segments record length segments total memory
(per channel)
2.5 ksample 5000 12.5 Msample
12.5 ksample 1000 12.5 Msample
125 ksample 100 12.5 Msample
Segmentation is active on all analog and logic channels and protocol
decoding.
Combinations with zoom and math functions are supported, but reduce the
effectively used number of segments
History mode function If active, the history mode always provides access to past acquisitions in the
segmented memory.
timestamp resolution 1.6 ps
time format relative, absolute
history player replays the recorded waveforms; start and stop waveform could be set;
repetition possible
R&S
®
RTH-K19
Advanced triggering
Additional trigger types:
TV/video: SDTV and HDTV broadcast standards; pattern, state, runt, slew rate, window, data2clock, serial pattern, timeout, interval
and protocol. For more details see Trigger system.