Parameters:
<State> ON | OFF
*RST: OFF
Example:
CALC:LIM:ACP:ALT3:ABS:STAT ON
Activates the absolute limit check for the third alternate channel.
CALCulate<1...2>:LIMit<1...2>:ACPower:ALTernate<1...11>:RESult?
This command queries the result of the limit check of the alternate channels.
To get a valid result, you have to perform a complete sweep with synchronization to
the sweep end before reading out the result. This is only possible in single sweep
mode.
This command is a query and therefore has no *RST value.
The suffix at CALCulate and LIMit are irrelevant for this command.
The suffix at ALTernate selects one of the 11 alternate channels.
Example:
CALC:LIM:ACP:ALT2:RES?
Queries the limit check results for the second alternate chan-
nels. The command would return, e.g. FAILED,FAILED Upper
and lower alternate channels violate a limit.
Usage: Query only
ACLR Result Query
The following commands query results specific to the adjacent channel power mea-
surement.
List of commands
"ACLR Result Query" on page 350
CALCulate<1...2>:MARKer<1...6>:FUNCtion:TXPower:RESult?
This command queries the total TX channel power. Use CALCulate<1...2>:
MARKer<1...6>:FUNCtion:POWer:RESult? on page 334 to query the power of
each TX channel.
The suffix at CALCulate and MARKer are irrelevant for this command.
Parameters:
<Return values> <Power>
The power is either the peak power or the channel power,
depending on the selected standard.
Example:
CALC:MARK:FUNC:TXP:RES?
Queries the total channel power over all TX channels.
Usage: Query only
Remote Control - Commands