Example:
CALC:MARK:FUNC:HARM ON
Activates the harmonic distortion measurement.
CALC:MARK:FUNC:HARM:NHAR 4
Sets the number of harmonics to 4.
CALC:MARK:FUNC:HARM:PRES
Adjusts the settings of the measurement.
CALC:MARK:FUNC:HARM:LIST?
CALC:MARK:FUNC:HARM:DIST? TOT
Queries the position of the harmonics and the total harmonic dis-
tortion.
Usage: Event
Measuring the AM Modulation Depth
The following commands configure AM Modulation Depth measurements. To perform
the actual measurement, use the commands described in section Chapter 13.6.2.4,
"Performing and Triggering Measurements", on page 300.
List of commands
CALCulate<1...2>:MARKer<1...6>:FUNCtion:MDEPth[:STATe]
on page 353
CALCulate<1...2>:MARKer<1...6>:FUNCtion:MDEPth:RESult?
on page 353
CALCulate<1...2>:MARKer<1...6>:FUNCtion:MDEPth[:STATe] <State>
This command turns the AM Modulation Depth measurement on and off.
The suffix at CALCulate and MARKer are irrelevant for this command.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
CALC:MARK:FUNC:MDEP ON
Activates the AM modulation depth measurement.
CALCulate<1...2>:MARKer<1...6>:FUNCtion:MDEPth:RESult?
This command queries the results of the AM modulation depth measurement.
The suffix at CALCulate and MARKer are irrelevant for this command.
Parameters:
<Return values> <ModulationDepth>
Modulation depth in %.
Example:
CALC:MARK:FUNC:MDEP:RES?
Queries the AM modulation depth.
Usage: Query only