Instructions

Version 11.00, May 2019
6 Rohde & Schwarz R&S
®
RTB Oscilloscope
Video trigger
trigger events
selectable line, all lines, even frame,
odd frame, all frames
supported standards
PAL, NTSC, SECAM, PAL-M, SDTV 576i,
HDTV 720p, HDTV 1080i, HDTV 1080p
sources
R&S
®
RTB2002
channel 1, channel 2, external trigger input
R&S
®
RTB2004
channel 1, channel 2, channel 3,
channel 4, external trigger input
sync pulse polarity
positive, negative
Pattern trigger
trigger events
logic condition between active channels
sources
R&S
®
RTB2002
channel 1, channel 2, logic channels from
D0 to D15 (with R&S
®
RTB-B1 option)
R&S
®
RTB2004
channel 1, channel 2, channel 3,
channel 4, logic channels from D0 to D15
(with R&S
®
RTB-B1 option)
state of channels
high, low, don’t care
logic between channels
and/or
condition
true, false
duration condition
smaller, greater, equal, unequal, inside
interval, outside interval, timeout
min. duration time
6.4 ns
max. duration time
13.5 s
Serial bus trigger
supported standards
R&S
®
RTB-K1 option
I
2
C/SPI (two- and three-wire)
R&S
®
RTB-K2 option
UART/RS-232/RS-422/RS-485
R&S
®
RTB-K3 option
CAN/LIN
Trigger sensitivity
R&S
®
RTB2002/R&S
®
RTB2004
with DC, AC, LF reject
input sensitivity > 5 mV/div
< 0.8 div (meas.)
2 mV/div ≤ input sensitivity < 5 mV/div
< 1.5 div (meas.)
input sensitivity < 2 mV/div
< 2 div (meas.)
with HF reject
all input sensitivities
< 1 div (meas.)
External trigger input
input impedance
R&S
®
RTB2002/R&S
®
RTB2004
1 MΩ ± 2 % with 9 pF ± 2 pF (meas.)
maximum input voltage at 1 mΩ
300 V (RMS), max. 400 V (V
p
),
derates at 20 dB/decade to 5 V (RMS)
above 250 kHz
trigger level
±5 V
sensitivity
300 mV (V
pp
)
input coupling
DC, AC, LF reject, HF reject
Trigger output (AUX OUT connector)
functionality
A pulse is generated for every acquisition
trigger event.
output voltage
at high impedance
0 V to 4.8 V
at 50 Ω
0 V to 2.4 V
pulse polarity
high active
output delay
depends on trigger settings