R&S®RTB2000 Oscilloscope Power of ten ❙ 70 MHz to 300 MHz ❙ 10-bit ADC ❙ 10 Msample standard memory ❙ 10.1" capacitive touchscreen Data sheet | Version 06.00 year RTB2000_dat_en_3607-4270-32_v0600.indd 1 06.05.
R&S®RTB2000 Oscilloscope At a glance Power of ten (10-bit ADC, 10 Msample memory and 10.1" touchscreen) combined with smart operating concepts make the R&S®RTB2000 oscilloscope the perfect tool for troubleshooting embedded designs, for university laboratories as well as for production and service departments. The largest display (10.1") with the highest resolution of its class (1280 × 800 pixel) works just like your smartphone.
The R&S®RTB2000 provides users with more than just an oscilloscope. It includes a logic analyzer, protocol analyzer, waveform and pattern generator and digital voltmeter. Dedicated operating modes for frequency analysis, mask tests and long data acquisitions are integrated. Debugging all kinds of electronic systems is easy and efficient – and satisfies the all-important rule of investment protection at a very attractive price.
See small signal details in the presence of large signals 10-bit A/D converter: uncovers even small signal details R&S®RTB2000 ❙❙ 10-bit vertical resolution Finest resolution for a 1 V signal 4 mV The R&S®RTB2000 features a customized Rohde & Schwarz designed 10-bit A/D converter that delivers a four-fold improvement compared to conventional 8-bit A/D converters. The increased resolution results in sharper waveforms with more signal details that would otherwise be missed.
Capture more time at full bandwidth ❙❙ 10 Msample standard, 20 Msample interleaved ❙❙ 160 Msample segmented memory with more than 13 000 recordings ❙❙ History mode: analysis of past acquisitions ❙❙ 1.25 Gsample/s, 2.5 Gsample/s interleaved 10 Msample standard and 20 Msample interleaved The R&S®RTB2000 offers a class-leading memory depth: 10 Msample per channel are available, even 20 Msample in interleaved mode. This is 10 times more than similar oscilloscopes in the same instrument class.
10.
n with gesture support 10.
X-in-1 oscilloscope Oscilloscope With a sampling rate of up to 2.5 Gsample/s and a memory depth of up to 20 Msample, the R&S®RTB2000 oscilloscope excels in its class. A waveform update rate of more than 50 000 waveforms/s ensures a responsive instrument that reliably catches signal faults. Included standard tools provide quick results, e.g. QuickMeas, mask tests, FFT, math, cursors and automatic measurements, including statistics.
Digital voltmeter The R&S®RTB2000 features a three-digit digital voltmeter (DVM) and six-digit frequency counter on each channel for simultaneous measurements. Measurement functions include DC, AC + DC (RMS) and AC (RMS). 1) 1) Included in scope of delivery. Frequency analysis mode Difficult-to-find faults often result from the interaction between time and frequency signals. The FFT function of the R&S®RTB2000 is activated at the push on a button and by entering center frequency and span.
Frequency response analysis (Bode plot) ❙❙ Analyze the frequency response of passive filters and amplifier circuits ❙❙ Perform control loop response measurements ❙❙ Perform power supply rejection ratio measurements ❙❙ Simple and fast documentation Perform low-frequency response analysis with an oscilloscope The R&S®RTB-K36 frequency response analysis (Bode plot) option lets you perform low-frequency response a nalysis on your oscilloscope easily and quickly.
Features and functionalities Amplitude profile The R&S®RTB-K36 frequency response analysis (Bode plot) option allows users to profile the amplitude output level of the generator. This helps to suppress the noise behavior of the DUT when performing a control loop response or power supply rejection ratio and to improve signal-to-noise ratio (SNR). It is possible to define up to 16 steps.
The best choice for education Professors especially like the password-protected education mode that disables automatic functions such as Autoset. This helps students understand the concepts. The built-in web server functionality enables professors to display their oscilloscope screen content to the classroom and over a network.
And there is so much more ... ❙❙ Efficient reporting capabilities ❙❙ Localized GUI and online help ❙❙ Fully upgradeable via software licenses ❙❙ Web server functionality for instrument access ❙❙ Extensive range of probes and accessories Grows with your needs The R&S®RTB2000 oscilloscopes flexibly adapt to needed project updates by installing software licenses. This applies to e.g. triggering and decoding of serial protocols and the history and segmented memory mode.
Oscilloscope portfolio R&S® RTH1000 RTC1000 RTB2000 RTM3000 RTA Bandwidth 60/100/200/350/500 MHz 1) 50/70/100/200/300 MHz 1) 70/100/200/300 MHz 1) 100/200/350/500 MHz/1 GHz 1) 200/3 Number of channels 2 plus DMM/4 2 2/4 2/4 4 Resolution 10 bit 8 bit 10 bit 10 bit 10 bit V/div 1 MΩ 2 mV to 100 V 1 mV to 10 V 1 mV to 5 V 500 µV to 10 V 500 µ V/div 50 Ω – 500 µV to 1 V 500 µ Vertical Horizontal Sampling rate per channel (in Gsample/s) 1.25 (4-channel model); 2.
RTA4000 RTE1000 RTO2000 RTP 200/350/500 MHz/1 GHz 1) 200/350/500 MHz/1/1.5/2 GHz 1) 600 MHz/1/2/3/4/6 GHz 1) 4/6/8/13/16 GHz 1) 4 2/4 2/4 (only 4 channels in 4 GHz and 6 GHz models) 4 10 bit 8 bit (up to 16 bit with HD mode) 8 bit (up to 16 bit with HD mode) 2) 8 bit (up to 16 bit with HD mode) 2) 500 µV to 10 V 500 µV to 10 V 1 mV to 10 V (500 μV to 10 V) 2) 500 µV to 1 V 500 µV to 1 V 1 mV to 1 V (500 μV to 1 V) 2) 1 mV to 1 V 2.
Base unit Vertical system Input channels Input impedance Analog bandwidth (–3 dB) Lower frequency limit (–3 dB) Analog bandwidth limits (max. –1.8 dB, min. –3.
Acquisition system Maximum realtime sampling rate Memory depth per channel Acquisition modes normal mode interleaved mode, following channels are not used simultaneously: channel 1 and channel 2 channel 3 and channel 4 logic channels normal If following channels are not used simultaneously: channel 1 and channel 2 channel 3 and channel 4 logic channels sample peak detect high resolution envelope average envelope + peak detect Number of averaged waveforms Waveform acquisition rate dot displa
Video trigger trigger events supported standards sources R&S®RTB2002 R&S®RTB2004 Pattern trigger sync pulse polarity trigger events sources R&S®RTB2002 R&S®RTB2004 state of channels logic between channels condition duration condition Serial bus trigger Trigger sensitivity R&S®RTB2002/R&S®RTB2004 External trigger input Trigger output (AUX OUT connector) min. duration time max.
Waveform measurements Automatic measurements measurements on channels, math waveforms, reference waveforms measurements on trigger signal reference levels statistics Cursor number of active measurements type functions Quick measurements function sources R&S®RTB2002 R&S®RTB2004 measurements displayed in diagram numerically displayed measurements burst width, count positive pulses, count negative pulses, count falling edges, count rising edges, mean value, RMS cycle, RMS, mean cycle, peak peak, peak+
Mask testing Sources R&S®RTB2002 R&S®RTB2004 Mask definition Result statistics Actions on mask violation channel 1, channel 2 channel 1, channel 2, channel 3, channel 4 acquired waveform with user-defined tolerance, can be stored and restored completed acquisitions, passed and failed acquisitions (absolute and in percent), test duration sound, acquisition stop, screenshot, save waveform, pulse out (AUX OUT connector) Waveform maths Number of math waveforms Functions Sources R&S®RTB2002 R&S®RTB2004 F
Display characteristics Diagram types XY mode manually changeable vertical window size Yt, XY, zoom, FFT parallel display of XY diagram and Yt diagrams of input signals for X, Y horizontal zoom with fast navigation, split screen with overview signal and zoomed signal split screen with Yt diagrams and dedicated frequency diagram sin(x)/x, linear, sample & hold lines, dots only 50 ms to 12.
Menu languages available menu languages: English German French Spanish Italian Portuguese Czech Polish Russian Simplified Chinese Traditional Chinese Korean Japanese online help, available languages: English undo/redo function Help Undo/redo Input and outputs Front Channel inputs External trigger input AUX OUT (BNC) Probe compensation output Pattern source (with R&S®RTB-B6 option only) Digital channel inputs Ground lug USB host interface trigger out reference frequency mask
General data Display Type Resolution Temperature Temperature loading 10.1" WXGA display with capacitive touch 1280 × 800 pixel (WXGA) operating temperature range storage temperature range Climatic loading Altitude Operating Nonoperating Mechanical resistance Vibration up to 3000 m above sea level up to 4600 m above sea level sinusoidal random Shock Maximum of sound pressure level EMC RF emission 5 Hz to 150 Hz, max. 1.8 g at 55 Hz; 0.
Options R&S®RTB-B1 Mixed signal option, additional 16 logic channels Vertical system Input channels Arrangement of input channels Input impedance Maximum input frequency Maximum input voltage Minimum input voltage swing Threshold groups Threshold level signal with minimum input voltage swing and hysteresis setting: normal hysteresis small hysteresis medium hysteresis large range predefined Threshold accuracy Comparator hysteresis Horizontal system Channel-to-channel skew Acquisition system Sampling rate
R&S®RTB-B6 Waveform generator and 4-bit pattern generator Waveform generator Resolution Sample rate Amplitude level high Z 50 Ω accuracy DC offset level high Z 50 Ω accuracy Sine frequency SFDR THD Rectangle frequency Pulse frequency edge time duty cycle Ramp, triangle, sinc, exponential frequency Arbitrary sample rate memory depth Noise bandwidth level Modulation AM function frequency depth FM function frequency deviation ASK function frequency ASK depth FSK function frequency FSK rate Sweep start frequenc
R&S®RTB-Bxx bandwidth upgrades Option Model R&S®RTB-B221 R&S®RTB-B222 R&S®RTB-B223 R&S®RTB-B241 R&S®RTB-B242 R&S®RTB-B243 R&S®RTB2002 R&S®RTB2002 R&S®RTB2002 R&S®RTB2004 R&S®RTB2004 R&S®RTB2004 Analog bandwidth upgrade from 70 MHz to 100 MHz 200 MHz 300 MHz 100 MHz 200 MHz 300 MHz R&S®RTB-K1 I2C triggering and decoding Bus configuration sources for SCL and SDA Trigger bit rate size of address size of data label list trigger events Decode offset for trigger on data data pattern width displayed sign
R&S®RTB-K2 UART/RS-232/RS-422/RS-485 triggering and decoding Bus configuration source for RX and TX bit rate Trigger Decode end of frame signal polarity data symbol size parity stop bits trigger events offset for trigger on data symbol data symbol pattern width displayed signals color coding of bus signal displayed format of data channel 1, channel 2, channel 3, channel 4, logic channels from D0 to D15 300/600/1200/2400/4800/9600/19200/ 38400/57600/115200 bps or user-selectable up to 3 Mbps timeout, no
Search search events frame event setup error event setup identifier setup data setup event table LIN triggering and decoding Bus configuration version bit rate Trigger polarity label list source trigger events identifier setup data setup Decode displayed signals color coding of bus signal displayed format of data frame table Search search events frame event setup error event setup identifier setup data setup event table 28 Rohde & Schwarz R&S®RTB Oscilloscope frame, error, identifier, identifi
R&S®RTB-K15 History and segmented memory Memory segmentation Fast-segmented mode History mode function additional memory segments for the acquisition number of segments 3 record segments total memory length (up to) (per channel) 10 ksample 13 107 131 Msample 20 ksample 13 107 262 Msample 50 ksample 4 369 218 Msample 100 2 621 262 Msample ksample 200 1 456 291 Msample ksample 500 624 312 Msample ksample 1 Msample 319 319 Msample 2 Msample 159 318 Msample 5 Msample 64 320 Msample 10 Msample 32 320 Msample
Ordering information Designation Type Choose your R&S®RTB2000 base model Oscilloscope, 70 MHz, 2 channels R&S®RTB2002 Oscilloscope, 70 MHz, 4 channels R&S®RTB2004 ® Base unit (including standard accessories: R&S RT-ZP03 passive probe per channel, power cord) Choose your bandwidth upgrade Upgrade of R&S®RTB2002 oscilloscopes to 100 MHz bandwidth R&S®RTB-B221 Upgrade of R&S®RTB2002 oscilloscopes to 200 MHz bandwidth R&S®RTB-B222 ® Upgrade of R&S RTB2002 oscilloscopes to 300 MHz bandwidth R&S®RTB-B223 Upgrade
Warranty Base unit All other items 4 Options Extended warranty, one year Extended warranty, two years Extended warranty with calibration coverage, one year Extended warranty with calibration coverage, two years Extended warranty with accredited calibration coverage, one year Extended warranty with accredited calibration coverage, two years 3 years 1 year R&S®WE1 R&S®WE2 R&S®CW1 R&S®CW2 R&S®AW1 R&S®AW2 Please contact your local Rohde & Schwarz sales office.
Service that adds value ❙ ❙ ❙ ❙ ❙ Worldwide Local and personalized Customized and flexible Uncompromising quality Long-term dependability Rohde & Schwarz The Rohde & Schwarz electronics group offers innovative solutions in the following business fields: test and measurement, broadcast and media, secure communications, cybersecurity, monitoring and network testing.