User manual

Measurements
R&S
®
RTB2000
108User Manual 1333.1611.02 ─ 06
Figure 7-1: Results of four active measurements
If a result cannot be determined, "?" is displayed. Adjust the horizontal and vertical set-
tings if the instrument cannot measure.
If the measurement result is outside the measurement range and clipping occurs, the
results are marked with "clipping+" or "clipping-". Adjust the vertical scale to get valid
results.
In addition to the current measurement results, you can enable a statistic evaluation. It
returns the current, minimum and maximum measurement values, the average and
standard deviation, and the number of measured waveforms. The results are shown in
a separate tab below the grid.
Figure 7-2: Statistic results of four active measurements
Remote commands are described in:
Chapter 15.7.2.2, "Measurements Results", on page 330
7.2.2 Measurement Types
The R&S RTB2000 provides many measurement types to measure time and amplitude
characteristics, and to count pulses and edges.
Automatic Measurements