Parameters:
<MeasType> FREQuency | PERiod | PEAK | UPEakvalue | LPEakvalue |
PPCount | NPCount | RECount | FECount | HIGH | LOW |
AMPLitude | MEAN | RMS | RTIMe | FTIMe | PDCYcle |
NDCYcle | PPWidth | NPWidth | CYCMean | CYCRms |
STDDev | CYCStddev | DELay | PHASe | BWIDth |
POVershoot | NOVershoot
See also: Chapter 7.2.2, "Measurement Types", on page 108.
FREQuency
Frequency of the signal. The result is based on the length of the
left-most signal period within the displayed section of the wave-
form of the selected channel.
PERiod
Length of the left-most signal period within the displayed section
of the waveform of the selected channel.
PEAK
Peak-to-peak value within the displayed section of the waveform
of the selected channel.
UPEakvalue
Maximum (upper) value within the displayed section of the
waveform of the selected channel.
LPEakvalue
Minimum (lower) value within the displayed section of the wave-
form of the selected channel.
PPCount | NPCount
Counts positive or negative pulses.
RECount | FECount
Counts the number of rising or falling edges.
HIGH | LOW
Mean value of the high or low level of a square wave.
AMPLitude
Amplitude of a square wave.
MEAN
Mean voltage value of the complete displayed waveform of the
selected channel.
RMS
RMS (root mean square) voltage value of the complete dis-
played waveform of the selected channel.
RTIMe | FTIMe
Rise or falling time of the left-most rising edge within the dis-
played section of the waveform of the selected channel. The ref-
erence level for this measurement is set with REFLevel:
RELative:MODE.
PDCycle | NDCycle
Positive or negative duty cycle.
PPWidth | NPWidth