User manual

Remote Control Commands
R&S
®
RTB2000
342User Manual 1333.1611.02 ─ 06
Suffix:
<m>
.
1, the suffix is irrelevant.
Return values:
<Value> Measurement result
Usage: Query only
CURSor<m>:XDELta[:VALue]?
Returns the time difference between two vertical cursor lines (Δt).
Suffix:
<m>
.
1, the suffix is irrelevant.
Return values:
<Delta> Range: -100E24 to 100E24
Increment: 0.1
Default unit: s
Usage: Query only
CURSor<m>:XDELta:INVerse?
Returns the inverse time difference between the two cursors (1/Δt, frequency).
Suffix:
<m>
.
1, the suffix is irrelevant.
Return values:
<DeltaInverse> Range: -100E24 to 100E24
Increment: 0.1
Default unit: 1/s
Usage: Query only
CURSor<m>:YDELta[:VALue]?
Queries the delta of the values in y-direction at the two cursor lines.
Suffix:
<m>
.
1, the suffix is irrelevant.
Return values:
<YDelta> Delta value
Usage: Query only
CURSor<m>:YDELta:SLOPe?
Returns the inverse value of the vertical difference (e.g. voltage difference) - the recip-
rocal of the vertical distance of two horizontal cursor lines: 1/
Δ
V.
Measurements