User manual

Fernsteuerbefehle
R&S
®
RTB2000
316Bedienhandbuch 1333.1611.03 ─ 05
Parameter:
<Polarity> POSitive | NEGative | EITHer
*RST: POS
SEARch:TRIGger:LEVel:RUNT:LOWer <LowerLevel>
Sets the lower voltage threshold for runt detection. A positive runt crosses the lower
level twice without crossing the upper level.
Parameter:
<LowerLevel> *RST: 400 mV
Std.-einheit: V
SEARch:TRIGger:LEVel:RUNT:UPPer
<UpperLevel>
Sets the upper voltage threshold for runt detection. A negative runt crosses the upper
level twice without crossing the lower level.
Parameter:
<UpperLevel> *RST: 600 mV
Std.-einheit: V
SEARch:TRIGger:RUNT:RANGe <Range>
Sets how the measured pulse width is compared with the given limit(s).
To set the width, use SEARch:TRIGger:RUNT:WIDTh.
To set the range ± Δt, use SEARch:TRIGger:RUNT:DELTa.
Parameter:
<Range> LONGer | SHORter | WITHin | OUTSide
LONGer
Finds pulses longer than the given width.
SHORter
Finds pulses shorter than the given width.
WITHin
Finds pulses inside the range width ± Δt.
OUTSide
Finds pulses outside the range width ± Δt.
*RST: LONG
SEARch:TRIGger:RUNT:WIDTh <Width>
Sets the reference runt pulse width, the nominal value for comparisons.
Messkurvenanalyse