User manual

Fernsteuerbefehle
R&S
®
RTB2000
341Bedienhandbuch 1333.1611.03 ─ 05
Suffix:
<m>
.
1..4
Selects the measurement place.
Abfrageparameter:
<PositivePeak> Maximum measurement result
Verwendung: Nur Abfrage
MEASurement<m>:RESult:WFMCount? <WaveformCount>
Returns the current number of measured waveforms.
The query for MIN and MAX values is not possible.
Suffix:
<m>
.
1..4
Selects the measurement place.
Abfrageparameter:
<WaveformCount> Number of measured waveforms
Verwendung: Nur Abfrage
MEASurement<m>:STATistics:WEIGht <AverageCount>
Sets the number of measured waveforms used for calculation of average and standard
deviation. Sets also the size of the statistics buffer.
Suffix:
<m>
.
1..4
Selects the measurement place.
Parameter:
<AverageCount> Bereich: 2 bis 1000
Inkrement: 1
*RST: 1000
MEASurement<m>:STATistics:VALue:ALL?
Returns all values from the statistics buffer.
Note: Valid buffered values can only be read if the acquisition is stopped. As long as
the acquisition is running, the buffer contents is changing and the buffered values are
not valid for reading.
Suffix:
<m>
.
1..4
Selects the measurement place.
Rückgabewerte:
<ValueList> Comma separated List of Values
Comma-separated list of statistical values
Messungen