CALibration
Calibration starts the self-alignment process. It can take several minutes. Consider
your timeout settings.
Calibration? returns information on the state of the self-alignment. Return values ≠
0 indicate an error.
Same as *CAL?.
Rückgabewerte:
<SelfAlignment> Numeric status indicator
CALibration:STATe?
Returns the overall state of the self-alignment.
Rückgabewerte:
<SelfAlignmentState> NOALignment | RUN | ERRor | OK | ABORt
NOALignment: no self-aligment was performed. Relevant for
service operations.
RUN: self-aligment is running
ERRor: an error occured.
OK: self-aligment has been performed successfully
ABORt: self-aligment has been cancelled
Verwendung: Nur Abfrage
SYSTem:NAME
Defines an instrument name.
Parameter:
<Name> String with max. 20 characters
SYSTem:DATE <Year>,<Month>,<Day>
Specifies the internal date for the instrument.
Parameter:
<Year> Inkrement: 1
Std.-einheit: a
<Month> Bereich: 1 bis 12
Inkrement: 1
<Day> Bereich: 1 bis 31
Inkrement: 1
Std.-einheit: d
Verwendung: SCPI-konform