Operating instructions

Getting Started R&S ESCI
1166.6004.12 5.2 E-1
Getting Started
The short and simple operating sequence given below permits fast putting into operation of the
instrument and setting of its basic functions. As a prerequisite, the IEC/IEEE-bus address, which is
factory-set to 20, must not have been changed.
1. Connect instrument and controller using IEC/IEEE-bus cable.
2. Write and start the following program on the controller:
CALL IBFIND("DEV1", receiver%) 'Open port to the instrument
CALL IBPAD(receiver%, 20) 'Inform controller about instrument address
CALL IBWRT(receiver%, '*RST;*CLS') ''Reset instrument
CALL IBWRT(receiver%, 'FREQ:CENT 100MHz') ' Set center frequency to 100 MHz
CALL IBWRT(receiver%, 'FREQ:SPAN 10MHz') ' Set span to 10 MHz
CALL IBWRT(receiver%, 'DISP:TRAC:Y:RLEV -10dBm')
' Set reference level to -10 dBm
The instrument now performs a sweep in the frequency range of 95 MHz to 105 MHz in analyzer mode.
3. To return to manual control, press the LOCAL key at the front panel