Operating instructions

R&S ESCI CALCulate:MARKer Subsystem
1166.6004.12 6.65 E-1
CALCulate<1|2>:MARKer<1 to 4>:FUNCtion:STRack:THReshold -330dBm to +30dBm
This command defines the threshold above which the largest signal is searched for in the selected
measurement window. The function is independent of the selected marker, i.e. the numeric suffix <1
to 4> of MARKer is irrelevant. It is only available in the frequency domain (span > 0)..
The response unit depends on the settings defined with CALC:UNIT.
Note: The entry of the search bandwidth is only possible if the Signal Track function is switched
on (CALC:MARK:FUNC:STR ON).
Example: "CALC:MARK:FUNC:STR:THR -50DBM" 'Sets the threshold for signal tracking
'in screen A to -50 dBm.
Characteristics: *RST value: -120 dBm
SCPI: device-specific
Mode: A-F
CALCulate<1|2>:MARKer<1 to 4>:FUNCtion:STRack:TRACe 1 to 3
This command defines the trace on which the largest signal is searched for in the selected
measurement window. The function is independent of the selected marker, i.e. the numeric suffix <1
to 4> of MARKer is irrelevant. It is only available in the frequency domain (span > 0).
Example: "CALC2:MARK:FUNC:STR:TRAC 3" 'Defines trace 3 in screen B as the
'trace for signal tracking.
Characteristics: *RST value: 1
SCPI: device-specific
Mode: A-F