Operating instructions

CALCulate:MARKer Subsystem R&S ESCI
1166.6004.12 6.68 E-1
CALCulate<1|2>:MARKer<1 to 4>:FUNCtion:SUMMary:PPEak:AVERage:RESult?
This command is used to query the result of the measurement of the averaged positive peak value in
the selected measurement window. The query is only possible if averaging has been activated
previously using CALCulate<1|2>:MARKer<1 to 4>:FUNCtion: SUMMary:AVERage.
The function is independent of the marker selection, i.e. the numeric suffix <1 to 4> in MARKer is
irrelevant. It is only available in the time domain (span = 0).
A complete sweep with synchronization to sweep end must be performed between switching on the
function and querying the measured value to obtain a valid query result. This is only possible in
single sweep mode.
Example: "INIT:CONT OFF" 'Switches to single-sweep mode.
"CALC:MARK:FUNC:SUMM:PPE ON" 'Switches on the function in screen A.
"CALC:MARK:FUNC:SUMM:AVER ON" 'Switches on the calculation of average
'in screen A.
"INIT;*WAI" 'Starts a sweep and waits for the end.
"CALC:MARK:FUNC:SUMM:PPE:AVER:RES?" 'Outputs the result of screen A.
Characteristics: *RST value: -
SCPI: device-specific
Mode: A-T
This command is only a query and therefore has no *RST value.
CALCulate<1|2>:MARKer<1 to 4>:FUNCtion:SUMMary:PPEak:PHOLd:RESult?
This command is used to query the result of the measurement of the positive peak value with active
peak hold function. The query is only possible if the peak hold function has been activated previously
using CALCulate<1|2>:MARKer<1 to 4>: FUNCtion:SUMMary:PHOLd.
The function is independent of the marker selection, i.e. the numeric suffix <1 to 4> of :MARKer is
irrelevant. It is only available in the time domain (span = 0).
A complete sweep with synchronization to sweep end must be performed between switching on the
function and querying the measured value to obtain a valid query result. This is only possible in
single sweep mode.
Example: "INIT:CONT OFF" 'Switches to single-sweep mode.
"CALC:MARK:FUNC:SUMM:PPE ON" 'Switches on the function in screen A.
"CALC:MARK:FUNC:SUMM:PHOL ON" 'Switches on the measurement of the
'peak value in screen A.
"INIT;*WAI" 'Starts a sweep and waits for the end.
"CALC:MARK:FUNC:SUMM:PPE:PHOL:RES?" 'Outputs the result of screen A.
Characteristics: *RST value: -
SCPI: device-specific
Mode: A-T
This command is only a query and therefore has no *RST value.