Operating instructions

CALCulate:MARKer Subsystem R&S ESCI
1166.6004.12 6.70 E-1
CALCulate<1|2>:MARKer<1 to 4>:FUNCtion:SUMMary:RMS:PHOLd:RESult?
This command queries the result of the measurement of the RMS value with active peak hold in the
selected measurement window. The query is only possible only if the peak hold function has been
activated previously using CALCulate<1|2>:MARKer<1 to 4>: FUNCtion:SUMMary:PHOLd.
The function is independent of the marker selection, i.e. the numeric suffix <1 to 4> of :MARKer is
irrelevant. It is only available in the time domain (span = 0).
A complete sweep with synchronization to sweep end must be performed between switching on the
function and querying the measured value to obtain a valid query result. This is only possible in
single sweep mode.
Example: "INIT:CONT OFF" 'Switches to single-sweep mode.
"CALC:MARK:FUNC:SUMM:RMS ON" 'Switches on the function in screen A.
"CALC:MARK:FUNC:SUMM:PHOL ON" 'Switches on the peak value
'measurement in screen A.
"INIT;*WAI" 'Starts a sweep and waits for the end.
"CALC:MARK:FUNC:SUMM:RMS:PHOL:RES?" 'Outputs the result of screen A.
Characteristics: *RST- value: -
SCPI: device-specific
Mode: A-T
This command is only a query and therefore has no *RST value.
CALCulate<1|2>:MARKer<1 to 4>:FUNCtion:SUMMary:MEAN[:STATe] ON | OFF
This command switches on or off the measurement of the mean value in the selected measurement
window.
The function is independent of the marker selection, i.e. the numeric suffix <1 to 4> of :MARKer is
irrelevant. It is only available in the time domain (span = 0)..
Note: The measurement is performed on the trace on which marker 1 is positioned. In order to
evaluate another trace, marker 1 must be positioned on another trace with
CALC:MARK:TRAC 1|2|3.
Example: "CALC:MARK:FUNC:SUMM:MEAN ON" 'Switches on the function in screen A.
Characteristics: *RST value: OFF
SCPI: device-specific
Mode: A-T
CALCulate<1|2>:MARKer<1 to 4>:FUNCtion:SUMMary:MEAN:RESult?
This command queries the result of the measurement of the mean value in the selected
measurement window. The function is independent of the marker selection, i.e. the numeric suffix
<1 to 4> of :MARKer is irrelevant. It is only available in the time domain (span = 0).
A complete sweep with synchronization to sweep end must be performed between switching on the
function and querying the measured value to obtain a valid query result. This is only possible in single
sweep mode.
Example: "INIT:CONT OFF" 'Switches to single-sweep mode.
"CALC:MARK:FUNC:SUMM:MEAN ON" 'Switches on the function in screen A.
"INIT;*WAI" 'Starts a sweep and waits for the end.
"CALC:MARK:FUNC:SUMM:MEAN:RES?"'Outputs the result of screen A.
Characteristics: *RST- value: -
SCPI: device-specific
Mode: A-T
This command is only a query and therefore has no *RST value.