Operating instructions

R&S ESCI CALCulate:MARKer Subsystem
1166.6004.12 6.71 E-1
CALCulate<1|2>:MARKer<1 to 4>:FUNCtion:SUMMary:MEAN:AVERage:RESult?
This command queries the result of the measurement of the averaged mean value in the selected
measurement window. The query is only possible if averaging has been activated previously using
CALCulate<1|2>:MARKer<1 to 4>:FUNCtion: SUMMary:AVERage.
The function is independent of the marker selection, i.e. the numeric suffix <1 to 4> of :MARKer is
irrelevant. It is only available in the time domain (span = 0).
A complete sweep with synchronization to sweep end must be performed between switching on the
function and querying the measured value to obtain a valid query result. This is only possible in
single sweep mode.
Example: "INIT:CONT OFF" 'Switches to single-sweep mode.
"CALC:MARK:FUNC:SUMM:MEAN ON" 'Switches on the function in screen A.
"CALC:MARK:FUNC:SUMM:AVER ON" 'Switches on the average value
'calculation in screen A.
"INIT;*WAI" 'Starts a sweep and waits for the end.
"CALC:MARK:FUNC:SUMM:MEAN:AVER:RES?" 'Outputs the result of screen A.
Characteristics: *RST- value: -
SCPI: device-specific
Mode: A-T
This command is only a query and therefore has no *RST value.
CALCulate<1|2>:MARKer<1 to 4>:FUNCtion:SUMMary:MEAN:PHOLd:RESult?
This command queries the result of the measurement of the mean value with active peak hold in the
selected measurement window. The query is only possible if the peak hold function has been
switched on previously using CALCulate<1|2>:MARKer<1 to 4>:FUNCtion:SUMMary:
PHOLd.
The query is possible only if the peak hold function is active. The function is independent of the
marker selection, i.e. the numeric suffix <1 to 4> of :MARKer is irrelevant. It is only available in the
time domain (span = 0).
A complete sweep with synchronization to sweep end must be performed between switching on the
function and querying the measured value to obtain a valid query result. This is only possible in
single sweep mode.
Example: "INIT:CONT OFF" 'Switches to single-sweep mode
"CALC:MARK:FUNC:SUMM:MEAN ON" 'Switches on the function in screen A
"CALC:MARK:FUNC:SUMM:PHOL ON" 'Switches on the peak value
'measurement in screen A
"INIT;*WAI" 'Starts a sweep and waits for the end
"CALC:MARK:FUNC:SUMM:MEAN:PHOL:RES?" 'Outputs the result of screen A.
Characteristics: *RST- value: -
SCPI: device-specific
Mode: A-T
This command is only a query and therefore has no *RST value.