Operating instructions

R&S ESCI CALCulate:MARKer Subsystem
1166.6004.12 6.73 E-1
CALCulate<1|2>:MARKer<1 to 4>:FUNCtion:SUMMary:SDEViation:PHOLd:RESult?
This command queries the maximum standard deviation value determined in several sweeps in the
selected measurement window. The query is possible only if the peak hold function is active.
The function is independent of the marker selection, i.e. the numeric suffix <1 to 4> of :MARKer is
irrelevant. It is only available in the time domain (span = 0).
A complete sweep with synchronization to sweep end must be performed between switching on the
function and querying the measured value to obtain a valid query result. This is only possible in
single sweep mode.
Example: "INIT:CONT OFF" 'Switches to single-sweep mode.
"CALC:MARK:FUNC:SUMM:SDEV ON" 'Switches on the function in screen A.
"CALC:MARK:FUNC:SUMM:PHOL ON" 'Switches on the peak value
'measurement in screen A.
"INIT;*WAI" 'Starts a sweep and waits for the end.
"CALC:MARK:FUNC:SUMM:SDEV:PHOL:RES?" 'Outputs the result of screen A.
Characteristics: *RST value: -
SCPI: device-specific
Mode: A-T
This command is only a query and therefore has no *RST value.
CALCulate<1|2>:MARKer<1 to 4>:FUNCtion:SUMMary:PHOLd ON | OFF
This command switches on or off the peak-hold function for the active time domain power
measurement in the indicated measurement window. The function is independent of the marker
selection, i.e. the numeric suffix <1 to 4> of :MARKer is irrelevant. It is only available in the time
domain (span = 0).
The peak-hold function is reset by switching it off and on again.
Example: "CALC:MARK:FUNC:SUMM:PHOL ON" 'Switches on the function in screen A.
Characteristics: *RST value: OFF
SCPI: device-specific
Mode: A-T
The peak-hold function is reset by switching off and on, again.
CALCulate<1|2>:MARKer<1 to 4>:FUNCtion:SUMMary:AVERage ON | OFF
This command switches on or off averaging for the active time domain power measurement in the
indicated window. The function is independent of the marker selection, i.e. the numeric suffix <1 to
4> of :MARKer is irrelevant. It is only available in the time domain (span = 0).
Averaging is reset by switching it off and on again.
The number of results required for the calculation of average is defined with
[SENSe<1|2>:]AVERage:COUNt.
It should be noted that synchronization to the end of averaging is only possible in single sweep
mode.
Example: "INIT:CONT OFF" 'Switches to single-sweep mode.
"CALC2:MARK:FUNC:SUMM:AVER ON" 'Switches on the calculation of average
'in screen B.
"AVER:COUN 200" 'Sets the measurement counter to 200.
"INIT;*WAI" 'Starts a sweep and waits for the end.
Characteristics: *RST value: OFF
SCPI: device-specific
Mode: A-T