CALCulate:PEAKsearch|PSEarch Subsystem
The CALCulate:PEAKsearch subsystem contains commands for the final measurement of scan results
in the test receiver mode.
Command PARAMETER Unit Comment
CALCulate<1|2>
:PEAKsearch
[:IMMediate]
:MARGin
:SUBRanges
:METHod
:PSEarch
[:IMMediate]
:MARGin
:SUBRanges
:METHod
<numeric_value>
<numeric_value>
SUBRange | PEAK
<numeric_value>
<numeric_value>
SUBRange | PEAK
dB
--
dB
--
:CALCulate<1|2>:PEAKsearch|PSEarch[:IMMediate]
This command generates a peak list from the scan results. For this purpose, a scan must be
performed prior to the command. The numeric suffix under CALCulate<1|2> is irrelevant.
Example: ":CALC:PEAK"
Characteristics: *RST value: -
SCPI: device specific
Mode: R
This command is an event and therefore has no *RST value and no query.
:CALCulate<1|2>:PEAKsearch|PSEarch:MARGin MINimun .. MAXimum
This command determines the margin (distance to the limit line) for generating the peak list. This
parameter is not used unless a limit line has been assigned to the trace. The margin, which can be
individually assigned to each limit line in the limit line menu, is not used when generating the peak
list.The numeric suffix under CALCulate<1|2> is irrelevant.
Example: ":CALC:PEAK:MARG 5 dB"
Characteristics: *RST value: 6 dB
SCPI: device specific
Mode: R
:CALCulate<1|2>:PEAKsearch|PSEarch:SUBRanges 1 to 500
This command determines the maximum number of subranges or peaks obtained when generating
the peak list. The numeric suffix under CALCulate<1|2> is irrelevant.
Example: ":CALC:PEAK:SUBR 10"
Characteristics: *RST value: 25
SCPI: device specific
Mode: R
CALCulate<1|2>:PEAKsearch|PSEarch:METHod SUBRange | PEAK
This command defines the method that is used to determine the maxima of a scan.The numeric
suffix in CALCULATE<1|2> is not significant.
Example: "CALC:PEAK:METH SUBR"
Characteristics: *RST value: PEAK
SCPI: device specific
Mode: R