Operating instructions

R&S ESCI CALCulate:STATistics Subsystem
1166.6004.12 6.79 E-1
CALCulate:STATistics Subsystem
The CALCulate:STATistics subsystem controls the statistical measurement functions in the instrument.
The measurement window cannot be selected with these functions. The numeric suffix in CALCulate is
therefore ignored.
COMMAND PARAMETERS UNIT COMMENT
CALCulate
:STATistics
:APD
[:STATe]
:CCDF
[:STATe]
:NSAMples
:SCALe
:AUTO
:X
:RLEVel
:RANGe
:Y
:UPPer
:LOWer
:PRESet
:RESult<1 to 3>?
<Boolean>
<Boolean>
<numeric_value>
ONCE
<numeric_value>
<numeric_value>
<numeric_value>
<numeric_value>
MEAN | PEAK | CFACtor | ALL
--
--
DBM
DB
query only
CALCulate:STATistics:APD[:STATe] ON | OFF
This command switches on or off the measurement of amplitude distribution (APD). On activating
this function, the CCDF measurement is switched off.
Example: "CALC:STAT:APD ON" 'Switches on the APD measurement.
Characteristics: *RST value: OFF
SCPI: device-specific
Mode: A
CALCulate:STATistics:CCDF[:STATe] ON | OFF
This command switches on or off the measurement of the complementary cumulative distribution
function (CCDF). On activating this function, the APD measurement is switched off.
Example: "CALC:STAT:CCDF ON" 'Switches on the CCDF measurement.
Characteristics: *RST value: OFF
SCPI: device-specific
Mode: A
CALCulate:STATistics:NSAMples 100 to 1E9
This command sets the number of measurement points to be acquired for the statistical
measurement functions.
Example: "CALC:STAT:NSAM 500" 'Sets the number of measurement points to be
'acquired to 500.
Characteristics: *RST value: 100000
SCPI: device-specific
Mode: A