Operating instructions

CALCulate:THReshold Subsystem R&S ESCI
1166.6004.12 6.82 E-1
CALCulate:THReshold Subsystem
The CALCulate:THReshold subsystem controls the threshold value for the maximum/minimum search
of markers. The measurement windows are selected by CALCulate 1 (screen A) or 2 (screen B).
COMMAND PARAMETERS UNIT COMMENT
CALCulate<1|2>
:DLINe<1|2>
:STATe
:THReshold
:STATe
:FLINe<1|2>
:STATe
:TLINe<1|2>
:STATe
<numeric_value>
<Boolean>
<numeric_value>
<Boolean>
<numeric_value>
<Boolean>
<numeric_value>
<Boolean>
DBM | DB |
DEG | RAD |
S | HZ | PCT
DBM | DB
HZ
S
CALCulate<1|2>:DLINe<1|2> MINimum .. MAXimum (depending on current unit)
This command defines the position of Display Line 1 or 2. These lines enable the user to mark any
levels in the diagram. The unit depends on the setting made with CALC:UNIT.
Example: "CALC:DLIN -20dBm"
Characteristics: *RST value: - (STATe to OFF)
SCPI: device-specific
Mode: R, A
CALCulate<1|2>:DLINe<1|2>:STATe ON | OFF
This command switches Display Line 1 or 2 (level lines) on or off.
Example: "CALC:DLIN2:STAT OFF"
Characteristics: *RST value: OFF
SCPI: device-specific
Mode: A
CALCulate<1|2>:THReshold MINimum to MAXimum (depending on current unit)
This command defines the threshold value for the maximum/minimum search of markers with
marker search functions MAX PEAK, NEXT PEAK, etc in the selected measurement window. The
associated display line is automatically switched on.
Example: "CALC:THR -82DBM" 'Sets the threshold value for screen A to -82 dBm.
Characteristics: *RST value: - (STATe to OFF)
SCPI: device-specific
Mode: R, A, FM