Operating instructions

R&S ESCI CALCulate:THReshold Subsystem
1166.6004.12 6.83 E-1
CALCulate<1|2>:THReshold:STATe ON | OFF
This command switches on or off the threshold line in the selected measurement window.The unit
depends on the setting performed with CALC:UNIT.
Example: "CALC2:THR:STAT ON" 'Switches on the threshold line in screen B.
Characteristics: *RST value: OFF
SCPI: device-specific
Mode: R, A, FM
CALCulate<1|2>:FLINe<1|2> 0...f
max
This command defines the position of the frequency lines.
The frequency lines mark the frequencies specified in the measurement window. Frequency lines are
only available with SPAN > 0.
Example: "CALC:FLIN2 120MHz"
Characteristics: *RST value: - (STATe to OFF)
SCPI: device-specific
Mode: R, A-F
CALCulate<1|2>:FLINe<1|2>:STATe ON | OFF
This command switches the frequency line on or off.
Example: "CALC:FLIN2:STAT ON"
Characteristics: *RST value: OFF
SCPI: device-specific
Mode: R, A-F
CALCulate<1|2>:TLINe<1|2> 0 ... 1000s
This command defines the position of the time lines.
The time lines mark the times specified in the measurement window. Time lines are only available
with SPAN = 0.
Example: "CALC:TLIN 10ms"
Characteristics: *RST value: - (STATe auf OFF)
SCPI: device-specific
Mode: A-Z
CALCulate<1|2>:TLINe<1|2>:STATe ON | OFF
This command switches the time line on or off.
Example: "CALC:TLIN2:STAT ON"
Characteristics: *RST value: OFF
SCPI: device-specific
Mode: A-Z