Operating instructions

CALibration Subsystem R&S ESCI
1166.6004.12 6.86 E-1
CALibration:RESult?
This command outputs the results of the correction data acquisition. The lines of the result table (see
section "Recording the correction data of R&S ESCI CAL key") are output as string data separated
by commas:
"Total Calibration Status: PASSED","Date (dd/mm/yyyy): 12/07/1999",
"Time: 16:24:54","Runtime:00.06"
Example: "CAL:RES?"
Characteristics: *RST value: --
SCPI: device-specific
Mode: all
CALibration:STATe ON | OFF
This command determines whether the current calibration data are taken into account by the
instrument (ON) or not (OFF).
Example: "CAL:STAT OFF" 'Sets up the instrument to ignore the calibration data.
Characteristics: *RST value: -
SCPI: conforming
Mode: all