Operating instructions

DIAGnostic Subsystem R&S ESCI
1166.6004.12 6.90 E-1
DIAGnostic<1|2>:SERVice:IQ:CALibration:PULSe:PRATe 10KHz | 62.5KHz | 80KHz | 100KHz |
102.4KHz | 200KHz | 500KHz | 1MHz |
2MHz | 4MHz
This command sets the frequency of the Pulse Cal signal.
The numeric suffix <1|2> has no meaning with this command.
Example: "DIAG:SERV:IQ:CAL:PULS:PRAT 80KHZ"
Characteristics: *RST value: 62.5 kHz
SCPI: device-specific
Mode: IQBB
DIAGnostic<1|2>:SERVice:IQ:INPut IQ | GND | CALDc | CALPulse
This command selects the baseband signal source.
IQ The female I and Q connectors of the FSQ are the baseband signal sources.
GND The baseband inputs are internally connected to ground.
CALDc The baseband signal source is the DC Cal signal. The voltage of this signal can be set with
diag:serv:iq:cal:dc.
CALPulse The baseband signal source is the Pulse Cal signal. The frequency of this signal can be set with
diag:serv:iq:prat.
The numeric suffix <1|2> has no meaning with this command.
Example: "DIAG:SERV:IQ:INP CALD"
Characteristics: *RST value: IQ
SCPI: device-specific
Mode: IQBB