Operating instructions

DISPlay Subsystem R&S ESCI
1166.6004.12 6.92 E-1
DISPlay:FORMat SINGle | SPLit
This command switches the measurement result display between FULL SCREEN and SPLIT
SCREEN. The coupling of settings between screen A and screen B can be selected with the
command INSTrument:COUPle.
In full-screen display the active measurement window can be selected with
DISPlay:WINDow<1|2>:SELect.
Example: "DISP:FORM SPL" 'Switches the display to 2 measurement windows.
Characteristics: *RST value: SINGle
SCPI: device-specific
Mode: R, A
DISPlay:ANNotation:FREQuency ON | OFF
This command switches the X axis annotation on or off.
Example: "DISP:ANN:FREQ OFF"
Characteristics: *RST value: ON
SCPI: conforming
Mode: all
DISPlay:LOGO ON | OFF
This command switches the company logo on the screen on or off.
Example: "DISP:LOGO OFF"
Characteristics: *RST value: ON
SCPI: device-specific
Mode: all
DISPlay:PSAVe[:STATe] ON | OFF
This command switches on or off the power-save mode of the display. With the power-save mode
activated the display including backlight is completely switched off after the elapse of the response
time (see command DISPlay:PSAVe:HOLDoff).
Note: This mode is recommended for preserving the display especially if the instrument is
exclusively operated via remote control.
Example: "DISP:PSAVe ON" 'Switches on the power-save mode.
Characteristics: *RST value: OFF
SCPI: device-specific
Mode: all
DISPlay:PSAVe:HOLDoff 1 to 60
This command sets the holdoff time for the power-save mode of the display. The available value
range is 1 to 60 minutes, the resolution 1 minute. The entry is dimensionless.
Example: "DISP:PSAV:HOLD 30"
Characteristics: *RST value: 15
SCPI: device-specific
Mode: all