Operating instructions

DISPlay Subsystem R&S ESCI
1166.6004.12 6.94 E-1
DISPlay:CMAP<1 to 33>:PDEFined BLACk | BLUE | BROWn | GREen | CYAN | RED | MAGenta |
YELLow | WHITe | DGRAy | LGRAy | LBLUe | LGREen | LCYan
| LRED | LMAGenta
This command defines the color table of the instrument using predefined color values. Each numeric
suffix of CMAP is assigned one or several graphical elements which can be modified by varying the
corresponding color setting.
The same assignment as for DISPlay:CMAP<1 to 33>:HSL applies.
Example: "DISP:CMAP2:PDEF GRE"
Characteristics: *RST value: --
SCPI: conforming
Mode: all
The values set are not changed by *RST.
DISPlay[:WINDow<1|2>]:SELect
This command selects the active measurement window. WINDow1 corresponds to SCREEN A,
WINDow2 to SCREEN B.
In FULL SCREEN mode, the measurements are only performed in the active measurement window.
Measurements are therefore initiated in the active window and result queries (marker, trace data and
other results) answered also in the active window.
Initiating measurements and queryĆ­ng results in the inactive window yields an error message
(execution error).
In split screen mode, the selection of the active window for result queries is irrelevant.
Note: - In FULL SCREEN mode, settings can also be performed in the inactive measurement
window. They become effective as soon as the corresponding window becomes
active.
Example: "DISP:WIND2:SEL 'Selects SCREEN B as active measurement window.
Characteristics: *RST value: SCREEN A active
SCPI: device-specific
Mode: A
This command is an event and therefore has no query.
DISPlay[:WINDow<1|2>]:SIZE LARGe | SMALl
This command switches the measurement window for channel and adjacent-channel power
measurements or for active FM demodulator (FS-K7) to full screen or half screen. Only "1" is allowed
as a numerical suffix.
Example: "DISP:WIND1:SIZE LARG"'Switches the measurement window to
full screen.
Characteristics: *RST value: SMALl
SCPI: device-specific
Mode: A, FM
DISPlay[:WINDow<1|2>]:TEXT[:DATA] <string>
This command defines a comment (max. 20 characters) which can be displayed on the screen in
the selected measurement window.
Example: "DISP:WIND2:TEXT 'Noise Measurement'"
'Defines the title for screen B
Characteristics: *RST value: "" (empty)
SCPI: conforming
Mode: all