Operating instructions

R&S ESCI DISPlay Subsystem
1166.6004.12 6.95 E-1
DISPlay[:WINDow<1|2>]:TEXT:STATe ON | OFF
This command switches on or off the display of the comment (screen title) in the selected
measurement window.
Example: "DISP:TEXT:STAT ON" 'Switches on the title of screen B.
Characteristics: *RST value: OFF
SCPI: conforming
Mode: all
DISPlay[:WINDow<1|2>]:TIME ON | OFF
This command switches on or off the screen display of date and time. The numeric suffix in
WINDow<1| 2> is irrelevant.
Example: "DISP:TIME ON"
Characteristics: *RST value: OFF
SCPI: device-specific
Mode: all
DISPlay[:WINDow<1|2>]:TRACE<1 to 3>:X:SPACing LINear | LOGarithmic
This command toggles between linear and logarithmic display in receiver mode
Example: "DISP:TRAC:X:SPAC LIN"
Characteristics: *RST value: LOGarithmic
SCPI: conforming
Modes: R, A
The numeric suffix in TRACE<1 to 3> is irrelevant.
DISPlay[:WINDow<1|2>]:TRACe<1 to 3>:Y[:SCALe] 10dB to 200dB
This command defines the display range of the Y axis (level axis) in the selected measurement
window with logarithmic scaling (DISP:TRAC:Y:SPAC LOG).
For linear scaling, (DISP:TRAC:Y:SPAC LIN | PERC) the display range is fixed and cannot be
modified. The numeric suffix in TRACe<1 to 3> is irrelevant.
Example: "DISP:TRAC:Y 110dB"
Characteristics: *RST value: 100dB
SCPI: device-specific
Mode: all
DISPlay[:WINDow<1|2>]:TRACe<1 to 3>:Y[:SCALe]:MODE ABSolute | RELative
This command defines the scale type of the Y axis (absolute or relative) in the selected
measurement window.
When SYSTem:DISPlay is set to OFF, this command has no immediate effect on the screen. The
numeric suffix in TRACe<1 to 3> is irrelevant.
Example: "DISP:TRAC:Y:MODE REL"
Characteristics: *RST value: ABS
SCPI: device-specific
Mode: all