Operating instructions

R&S ESCI DISPlay Subsystem
1166.6004.12 6.97 E-1
DISPlay[:WINDow<1|2>]:TRACe<1 to 3>:Y[:SCALe]:RPOSition 0 to 100PCT
This command defines the position of the reference value in the selected measurement window.
The numeric suffix in TRACe<1 to 3> is irrelevant.
In operating mode NETWORK (Tracking Generator /Ext. Generator Option R&S FSP-B9/B10) with
active normalization, RPOSition defines the reference point for the output of the normalized
measurement results.
With the FM demodulator (FS-K7) switched on, the reference position for result display RF POWER
and SPECTRUM is kept separate from that for result display FM.
Example: "DISP:TRAC:Y:RPOS 50PCT"
Characteristics: *RST value: 100PCT(SPECTRUM mode)
50 PCT (NETWORK and FM DEMOD mode)
50 PCT (Mode FM-DEMOD mit FM display)
100 PCT(Mode FM-DEMOD mit AF spectrum display of FM)
SCPI: conforming
Mode: A
DISPlay[:WINDow<1|2>]:TRACE<1 to 3>:Y[:SCALe]:PDIVision <numeric_value>
This command defines the scaling of the Y axis in the current unit.
Example: "DISP:TRAC:Y:PDIV +1.20"
Characteristics: *RST value: -
SCPI: conforming
Mode: FM
The numeric suffix in TRACE<1 to 3> is irrelevant. The command is only available with option R&S
FS-K7 (FM Demodulator).
DISPlay[:WINDow<1|2>]:TRACe<1 to 3>:Y[:SCALe]:BOTTom <numeric_value>
This command defines the minimum grid level in the current unit for the scan display in the receiver
mode.
Example: "DISP:TRAC:Y:BOTT -20" The minimum grid level is set to -20 dBuV
(pre-condition: the default unit has not been
changed)
Characteristics: *RST value: 0
SCPI: conforming
Mode: R
The numeric suffix in TRACE<1 to 3> is irrelevant.
DISPlay[:WINDow<1|2>]:TRACe<1 to 3>:Y:SPACing LINear | LOGarithmic| LDB
This command toggles between linear and logarithmic display in the selected measurement
window.On a linear scale, switchover between the unit % (command DISP:WIND:TRAC:Y:SPAC
LIN) and the unit dB (command DISP:WIND:TRAC:Y:SPAC LDB) is also possible.
If the FM demodulator (FS-K7) is active and result display AF spectrum of FM is selected, only the
parameters LINear and LOGarithmic are permissible.
The numeric suffix in TRACe<1 to 3> is irrelevant.
Example: "DISP:TRAC:Y:SPAC LIN"
Characteristics: *RST value: LOGarithmic
SCPI: conforming
Mode: A, FM