Operating instructions

DISPlay Subsystem R&S ESCI
1166.6004.12 6.98 E-1
DISPlay[:WINDow<1|2>]:TRACe<1 to 3>:MODE WRITe | VIEW | AVERage | MAXHold | MINHold
This command defines the type of display and the evaluation of the traces in the selected
measurement window. WRITE corresponds to the Clr/Write mode of manual operation. The trace is
switched off (= BLANK in manual operation) with DISP:WIND:TRAC:STAT OFF.
The number of measurements for AVERage, MAXHold and MINHold is defined with the command
SENSe:AVERage:COUNt or SENSe:SWEep:COUNt. It should be noted that synchronization to the
end of the indicated number of measurements is only possible in single sweep mode.
If calculation of average values is active, selection between logarithmic and linear averaging is
possible. For more detail see command SENSe:AVERage:TYPE.
Example: "SWE:CONT OFF" Switching to single-sweep mode.
"SWE:COUN 16" 'Sets the number of measurements to 16.
"DISP:WIND1:TRAC3:MODE MAXH" 'Switches on the calculation of the for
'trace 3 in screen A.
"INIT;*WAI" 'Starts the measurement and waits for the end of the
'16 sweeps.
Characteristics: *RST value: WRITe for TRACe1, STATe OFF for TRACe2/3
SCPI: device-specific
Mode: all
DISPlay[:WINDow<1|2>]:TRACE<1 to 3>[:STATe] ON | OFF
This command switches on or off the display of the corresponding trace in the selected
measurement window.
Example: "DISP:WIND1:TRAC3 ON"
Characteristics: *RST value: ON for TRACe1, OFF for TRACe2 to 4
SCPI: conforming
Mode: all
DISPlay[:WINDow<1|2>]:TRACE<1 to 3>:SYMBol CROSs | OFF
This command switches on or off the indication of the peak list or the final measurement result on
the trace.
CROSs The single value is indicated as x.
OFF No indication.
Example: "DISP:TRAC:SYMB CROS"
Characteristics: *RST value: OFF
SCPI: device-specific
Mode: R
:DISPlay:BARGraph:LEVel:LOWer?
This command queries the minimum level of the bargraph.
Example: ":DISP:BARG:LEV:LOW?"
Characteristics: *RST value: --
SCPI: device-specific
Mode: R