Operating instructions

INSTrument Subsystem R&S ESCI
1166.6004.12 6.112 E-1
INSTrument Subsystem
The INSTrument subsystem selects the operating mode of the unit either via text parameters or fixed
numbers.
COMMAND PARAMETERS UNIT COMMENT
INSTrument
[:SELect]
:NSELect
:CONFigure<1|2>
[:DISPlay]
:MIXed
SANalyzer| ADEMod | RECeiver |
<numeric_value>
SANalyzer | RECeiver
<Boolean>
:COUPle
:ATTenuation
:BANDwidth
:BWIDth
:CENTer
:DEModulation
:GAIN
:PRESelector
:PROTection
:SPAN
NONE | RLEVel | CF_B | CF_A
ALL | NONE
ALL | NONE
ALL | NONE
ALL | NONE
ALL | NONE
ALL | NONE
ALL | NONE
ALL | NONE
ALL | NONE
no query
INSTrument[:SELect] SANalyzer | RECeiver | ADEMod
This command switches between the operating modes by means of text parameters.
Parameter: SANalyzer: Spectrum analysis
ADEMod FM demodulator
RECeiver: EMI receiver
Example: "INST SAN" 'Switches the instrument to SPECTRUM.
Characteristics: *RST value: RECeiver
SCPI: conforming
Mode: all
Changeover to ADEMod is only possible with FM Demodulator FS-K7 option installed.
INSTrument:NSELect 1 | 3 | 6
This command switches between the operating modes by means of numbers.
Parameter: 1: Spectrum analysis
3: FM demodulator
6: EMI Receiver
Example: "INST:NSEL 1" 'Switches the instrument to SPECTRUM.
Characteristics: *RST value: 6
SCPI: conforming
Mode: all
Changeover to 3 is only possible with FM Demodulator FS-K7 option installed.