Operating instructions

R&S ESCI INSTrument Subsystem
1166.6004.12 6.115 E-1
INSTrument:COUPle:PRESelector ALL | NONE
This command couples the activation of preselection for the analyzer and receiver modes to each
other.
Preselection cannot be switched off in the receiver mode. In practice, this means that preselection is
always switched on automatically when changing to the analyzer mode with coupling activated.
Example: "INST:COUP:PRES ALL"
Characteristics: *RST value: ALL
SCPI: device specific
Mode: R, A
INSTrument:COUPle:PROTection ALL | NONE
This command couples the RF safety attenuation of 10 dB for the analyzer and receiver modes to
each other.
To prevent damage to the RF input, switch on this coupling.
Example: "INST:COUP:PROT ALL"
Characteristics: *RST value: ALL
SCPI: device specific
Mode: R, A
INSTrument:COUPle:SPAN ALL | NONE
This command couples the start and stop frequencies of sweep and scan for the analyzer and
receiver modes to each other.
Example: "INST:COUP:SPAN ALL"
Characteristics: *RST value: NONE
SCPI: device specific
Mode: R, A