MMEMory:STORe<1|2>:TRACe 1 to 3,<file_name>
This command stores the selected trace (1 to 3) in the measurement window indicated by
STORe<1|2> (screen A or B) in a file with ASCII format. The file format is described in chapter 4 in
the TRACE menu under the ASCII-FILE EXPORT softkey.
The decimal separator (decimal point or comma) for floating-point numerals contained in the file is
defined with the command FORMat:DEXPort:DSEParator.
The file name includes indication of the path and the drive name. Indication of the path complies with
DOS conventions.
Parameter: 1 to 3 := selected measurement curve Trace 1 to 3
<file_name> := DOS file name
Example: "MMEM:STOR2:TRAC 3,'A:\TEST.ASC'"
'Stores trace 3 from screen B in the file
'TEST.ASC on a disk.
Characteristics: *RST value: -
SCPI: device-specific
Mode: all
This command is an event and therefore has no *RST value and no query.
:MMEMory:STORe:FINal <file_name>
This command stores the final measurement data in an ASCII file.
The file name includes indication of the path and may also include the drive. The path name
corresponds to the DOS conventions.
Parameter: <file_name> := DOS file name
Example: ":MMEM:STOR:FIN 'A:\TEST.ASC'"
Characteristics: *RST value: -
SCPI: device-specific
Mode: R
This command is an event and therefore has no *RST value and no query.
MMEMory:CLEar:STATe 1,<file_name>
This command deletes the instrument setting selected by <file_name>. All associated files on the mass
memory storage are cleared. A list of the extensions used is included under MMEMory:LOAD:STATe.
The file name includes indication of the path and may also include the drive. The path name
complies with DOS conventions.
Parameter: <file_name> ::= DOS file name without extension
Example: "MMEM:CLE:STAT 1,'TEST'"
Characteristics: *RST value: -
SCPI: device-specific
Mode: all